| Model Number |
0064599 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional serial numbers continued.Additional lot numbers continued.
|
| |
|
Event or Problem Description
|
|
A review of the events indicated that three (3) patient samples tested on the galileo neo, automated blood bank system produced inaccurate aborh results.In two (2) tests, the instrument incorrectly produced a negative result for anti-d.In a third test; an a positive cord blood sample mistyped as ab positive.In each circumstance, subsequent testing of relevant reagent retention lots produced accurate results.Additional investigations related to quality control processes; test procedures and other facts could not determine any direct cause of the missed antibody screening and the malfunctions are allocated against the analytical instrument.
|
| |
|