| Model Number |
72213N |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event is unknown; awareness date has been used for this field.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd ultra-fine¿ micro pen needles 32g x 6mm (100 count) the pen needles couldn't prime.This occurred 5 times.There was no report of patient impact.The following information was provided by the initial reporter: consumer reported found 5 pen needles clog during the flow check.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.A lot history review was carried out and no related non conformances were raised in association with this packaged lot concluding all inspections were performed as per the applicable operations and met qc specifications.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd ultra-fine¿ micro pen needles 32g x 6mm (100 count) the pen needles couldn't prime.This occurred 5 times.There was no report of patient impact.The following information was provided by the initial reporter: consumer reported found 5 pen needles clog during the flow check.
|
| |
|