U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Disturbances (2140); Halo (2227)
Date of Event 09/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Weight: unknown/ information asked but not available.Ethnicity: unknown/ information asked but not available.Device evaluation: the product testing could not be performed as the product was not returned for evaluation.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing records review was performed, and no nonconformity report was found as part of this manufacturing records review.The product was manufactured and released according to specifications.A search in complaint system revealed that no other complaints were received for this production order number.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be verified.An attempt was made to obtain the missing information; however, the information was not available.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted from a patient's right eye due to mechanical failure.The account clarified that by mechanical failure, they meant that halos and shadows around vision during the day and evening making it difficult to drive.It was confirmed that the patient did not have any pre-existing issue which contributed to the reported issues.The incision was enlarged, and the lens was taken out in whole.No other surgical interventions were required.Another johnson and johnson lens of a different model and diopter (aab00 +19.50) was used as a replacement.The patient is stable, and no injury was reported.No further information was provided.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: nov 3, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received wrapped in gauze.No additional materials were received.Visual inspection under magnification revealed that the complaint lens was received coated in viscoelastic residue.The lens was cleaned and, no issues that could cause or contribute to the complaint issues could be identified.The complaint issues were not confirmed.No additional issues were observed during evaluation.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15705333
Report Number3012236936-2022-02722
Device Sequence Number14738694
Product Code MFK
UDI-Device Identifier05050474750654
UDI-Public(01)05050474750654(17)240916
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/16/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU190
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date10/31/2022
Supplement Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AMVISC VISCOELASTIC.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-