Model Number 71436128
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, while a trial reduction during uni procedure, a journey uni tib tr ins sz 1-2/8mm piece broke from the main body and was then retrieved by forceps.It is unknown how the surgery was completed.The procedure was finished with a non-significant delay.No further complications were reported.
 
Additional Manufacturer Narrative
Internal reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation but the pictures were reviewed, and revealed that the trial insert broke.The clinical/medical investigation concluded that the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.Based on a review of the information provided, the broke piece was retrieved with forceps.According to the report, it is unknown how the surgery was completed.However, it was reported the surgery was completed with a non-significant delay.Since there was no harm alleged to this patient and no other complications were reported, no further clinical/medical assessment is warranted at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable, damage may occur during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported incident could be corroborated as it may have occurred due to its extensive use.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY UNI TIB TR INS SZ 1-2/8MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15705353
Report Number1020279-2022-04613
Device Sequence Number5145281
Product Code HSX
UDI-Device Identifier03596010588067
UDI-Public03596010588067
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K113038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71436128
Device Catalogue Number71436128
Device Lot Number20EM00148
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2022
Supplement Date Received by Manufacturer11/24/2022
11/25/2022
Initial Report FDA Received Date10/31/2022
Supplement Report FDA Received Date11/28/2022
11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/01/2020
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-