MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
APB-2-2-HX-ES |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information that an axium coil detached prematurely.The coil became pre-detachable in the microcatheter.Removed and treated with another coil.The coil was removed from the patient by withdrawing the catheter, no surgical or medical intervention was required.The pushwire was not bent or broken.There was no friction or difficulty during delivery.The physician did not reposition the coil.The physician did not attempt to detach the coil.The physician did not rotate the delivery pusher during the procedure.Continuous flush was administered during the procedure.The devices were prepared as indicated in the instructions for use (ifu).The patient was being treated for an unruptured, saccular aneurysm in the a com location.The mix diameter was 4mm and the neck diameter was 1.5mm.The patient's blood flow was normal and vessel tortuosity was moderate.No patient symptoms or complications were reported. ancillary devices: fubuki 8f guide catheter, headway17 microcatheter, synchro14 guidewire.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the axium prime coil (lot# 223153479) found the coil was returned without introducer sheath.The pushwire and release wire were found to be broken at proximal end.The pushwire was found to be kinked at ~28.0cm, at ~152.4cm and wavy at ~168.8cm for ~5.0cm from broken end.This is indicative of detachment attempts by manual methods.The coin was found to be still against the lumen stop.The shield coil was found to be intact.The implant coil was already detached and not returned.No other anomalies were observed.Under the microscope, the outer jacket was then removed to gain access to the coin.The coin appeared to be in good condition.The coin was measured to be 0.074mm at 0.063mm which is within specification, 0.089mm at 0.127mm, and 0.092mm at 0.275mm which were within specification.The lumen stop appeared to be visually acceptable.The retainer ring was found to be visually acceptable and measured to be 0.0050¿ which is within specification.Based on the analysis performed, the customers report of ¿premature detachment¿ was confirmed as the axium prime was returned with the implant already detached; however, the cause could not be determined.Since the implant coil and included detach, element were not returned for analysis, any contributing factors could not be determined.A possible cause for premature detachment is the coil not being retracted in a one-to-one motion with the implant pusher during repositioning.The damages (broken/kinked) to the pushwire are consistent with manual detachment.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|