|
We checked the returned unit and confirmed that the nozzle gluing missing.Based on the result, we concluded that it was caused due to the physical damage applied on the nozzle gluing.In addition, we confirmed that the bending rubber leak, and the bending rubber cut; however, they are not the main cause, and/or irrelevant to the alleged complaint.In terms of gluing missing, the possibility of dropping into human body could not be denied.Moreover, based on the technical report "hr-rpt-0585 (nozzle)" and/or the risk analysis results, it was evaluated to submit mdr.
|