|
|
| Model Number |
BB8G55R15 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
10/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information that during use of a custom tubing pack, the customer reported a leak was noticed from the line connection.The customer stated that the blood loss was minimal, and a blood transfusion was not required.The device was used to complete the procedure.There was no additional patient impact associated with this event.Medtronic received additional information that there was no visible air in system/tubing reported.The same leak did not appear in multiple packs.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual inspection shows an additional tie band installed on one of the "y" fitting connections.The additional tie band was removed.Pressure integrity testing was performed at 3 l/pm with 23 psi, (1189 mmhg) of back pressure for 10 minutes.During the pressure integrity testing there was a leak observed from the "y" fitting/tubing interface.Reason for return was confirmed.Conclusion: complaint is confirmed for ¿leaks¿.After evaluation the cause of this complaint could not be determined.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.Complaints received from august 2022 through november 2022 for the same failure mode were reviewed and showed no trends warranting escalation related to this occurrence.There were no adverse patient effects as a result of this incidence.Medtronic will continue to monitor for future occurrences and trends.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|