|
(b)(4).Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the pxw35 device was returned with no apparent damage, fully functional.When tested for functionality the device fired and formed the remaining 25 staples as intended.The device fired without any difficulties and the staples were noted to have the proper box shape.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: with two tissue forceps, pick up each wound edge individually and approximate the edges.Or, with one tissue forceps, pull skin edges together until edges evert.Or apply tension to either end of the incision, such that the tissue edges begin to approximate themselves.One forceps can be used to ensure that the edges are everted.Position the instrument over the everted skin edges, aligning the instrument arrow with the center of the incision at a 50° - 60° angle from the skin.Squeeze the trigger until you touch plastic trigger to plastic handle (plastic to plastic), then release the trigger and move the instrument off the incision.The event described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|