| Model Number |
123612A |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter was inserted on (b)(6) 2022.On (b)(6) 2022, patient informed doctor about the leaking and ordered to change the urinary foley catheter.When aspirating water from the balloon it was found that the balloon of the catheter had ruptured as the backflow from the balloon was yellow color like the urine color.Nurse was able to remove the catheter from the patient and reinserted a new foley catheter.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter was inserted on (b)(6) 2022.On (b)(6) 2022, patient informed doctor about the leaking and ordered to change the urinary foley catheter.When aspirating water from the balloon it was found that the balloon of the catheter had ruptured as the backflow from the balloon was yellow color like the urine color.Nurse was able to remove the catheter from the patient and reinserted a new foley catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed and the cause was unknown.A potential root cause for this potential failure mode could be due to short latex dwell time or latex dip speed out too slow.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "to deflate catheter balloon.Gently, insert a syringe in the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If the tails, contact an adequately trained professional for assistance as detected by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter was inserted on (b)(6)2022.On (b)(6)2022, patient informed doctor about the leaking and ordered to change the urinary foley catheter.When aspirating water from the balloon it was found that the balloon of the catheter had ruptured as the backflow from the balloon was yellow color like the urine color.Nurse was able to remove the catheter from the patient and reinserted a new foley catheter.
|
| |
|