| Model Number |
DI-300 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the device had a leakage problem.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Health impact and evaluation codes: updated a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing was performed.Visual inspection found no damage / broken and contamination was observed.Functional testing found a leak is observed between tube with the adapter tube in the units received.By the physical evaluation of the unit, the complaint is confirmed.The root cause of the reported issue was found to be lack of adhesive between tube with female luer and adapter tube during bonding assembly.An awareness notification was conducted by the quality engineer on (b)(6) 2021 to the production personnel to explain the importance of adherence in the procedure.Action taken after the manufactured lot reported in this complaint this remediation mdr was generated under protocol (b)(6) as a result of warning letter cms# (b)(4).
|
| |
|