• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC. SHOCKPULSE LITHOTRIPSY TRANSDUCER.; LITHOTRIPTOR, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. SHOCKPULSE LITHOTRIPSY TRANSDUCER.; LITHOTRIPTOR, ULTRASONIC Back to Search Results
Model Number SPL-T
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to olympus while preparing for use, a transducer error code and other error messages displayed when the shockpulse lithotripsy transducer was plugged in.There were no reports of patient harm associated with this event.
 
Additional Manufacturer Narrative
The device has not been received to date.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.Based on the results of the investigation, since the device was not returned for evaluation, a root cause of the reported event could not be determined.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿regarding check probe illumination, possible causes may be ¿the probe is loose¿ or ¿the probe has failed or has a crack.¿ troubleshooting this problem includes ¿remove the probe and clean the mating surfaces of the probe and transducer.Reattach probe using the wrench,¿ or ¿replace probe.¿ " ¿for the problem with error indicator illumination, ¿the transducer may have malfunctioned.¿ troubleshooting this problem includes ¿reboot generator by turning power off and then back on.Plug in a second transducer to the generator.¿ in addition, the ifu advises ¿it is always recommended that a spare transducer and probe assembly be available.¿ ¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE LITHOTRIPSY TRANSDUCER.
Common Device Name
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
9600 louisiana avenue north
brooklyn park, MN 55445
5082077661
MDR Report Key15709725
Report Number3011050570-2022-00120
Device Sequence Number5144895
Product Code FEO
UDI-Device Identifier00855279005023
UDI-Public00855279005023
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K142428
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSPL-T
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2022
Supplement Date Received by Manufacturer11/28/2022
Initial Report FDA Received Date11/01/2022
Supplement Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/25/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-