NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number P10.7-80E245
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Codes Failure of Implant (1924); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received that at a follow-up appointment, the regenerate was noticed to have collapsed back to 16mm.X-ray imaging revealed what appeared to be a crown fracture.The nail has been explanted and exchanged for a new one.
 
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
No additional information has been provided.
 
Additional Manufacturer Narrative
Additional data: b5, d9, h3, h6, h10.Device evaluation: visual inspection of the returned precice nail revealed the housing tube had cracks at the anti-rotation lug in the crown region.X-ray images of the internal components confirmed a broken anti-rotation lug and no other internal damage was found.The nail has been functionally tested and was able to be distracted and retracted with the high-speed magnet for stroke testing.The maximum and minimum lengths were measured at 243.76mm and 325.86mm, respectively, giving a total stroke measurement of 82.1, which meets the specification.Distraction force testing was measured at 280.1 lbs.Which meets the specification.The work order was reviewed and confirmed the device passed all inspections per the acceptance tests.The x-ray images taken at the time of manufacturing indicate the anti-rotation lug (crown) was seated properly in the housing body.The inspection data for the lots of anti-rotation lug and housing tube have been reviewed and confirmed the parts met design specifications as per the engineering drawings.Based on the provided x-ray images of the patient¿s body, the reported failure of loss of distraction was confirmed.The loss of distraction is likely due to the broken anti-rotation lug.Although the exact cause of broken nail/broken anti-rotation lug is unknown, based on the type of breakage observed it is likely the nail broke due to stress generated from weight bearing activities.Device labeling: per the precice instructions for use (ifu), ¿the precice intramedullary limb lengthening nail cannot withstand the stress of full weight bearing for tibia and femur applications.For humerus applications, patients should not bear any weight on the treated limb.Patients should utilize external support and/or restrict activities until consolidation occurs¿.Device records review: a review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE SYSTEM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr, ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key15709980
Report Number3006179046-2022-00270
Device Sequence Number12048299
Product Code HSB
UDI-Device Identifier00812258020828
UDI-Public812258020828
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K191336
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberP10.7-80E245
Device Lot Number0081022
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2022
Supplement Date Received by Manufacturer12/02/2022
Initial Report FDA Received Date11/01/2022
Supplement Report FDA Received Date12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight65 KG
-
-