ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION UV ABSORBING IOL; INTRAOCULAR LENS Back to Search Results
Model Number DAT015
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Impairment (2138)
Date of Event 02/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported that following an intraocular lens (iol) implant procedure, the patient experienced blurry vision.The clinical reason was refractive error.The iol was explanted and exchanged in a secondary procedure for a monofocal lens.Additional information has been requested and received stated that patient had blurred vision at all distances and had the trouble with adaptation to the multi focal lens.There was no patient harm.The surgeon prognosis was good, the patient was not hospitalized and the issue has been resolved.
 
Additional Manufacturer Narrative
The account indicated the use of a qualified associated cartridge and viscoelastic.The handpiece used if only qualified for use with company lens models.Each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the company lens, 23.0 diopter, (1.5 extended depth of focus) was replaced with a non-company, 22.0 diopter, monofocal lens model.The product was not returned for evaluation.Additional information was provided that, in the surgeon's opinion, the product did not cause or contribute to the event.It was reported that the event was due to a refractive error and that the patient had "trouble with adaptation to the multifocal." no additional information has been provided at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY EXTENDED VISION UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15710055
Report Number1119421-2022-02327
Device Sequence Number12417176
Product Code POE
UDI-Device Identifier00380652424691
UDI-Public00380652424691
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/28/2025
Device Model NumberDAT015
Device Lot Number15103049
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/06/2022
Supplement Date Received by Manufacturer12/13/2022
Initial Report FDA Received Date11/01/2022
Supplement Report FDA Received Date01/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR.; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D.; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-