| Model Number |
MAF-GM |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation and the customers allegation was able to be confirmed.During the device evaluation it was observed that the battery could not be removed due to damage of the card slot caused by external factors.It was also observed that the coating of the light guide serpentine was peeling due to buckling.It was observed that there were stains on the light guide bundle due to external factors.It was also observed that the curved rubber adhesive was missing.Also, the watertightness was not maintained due to holes in the curved rubber caused by external factors.It was observed that the watertightness of the forceps channel was not maintained due perforation caused by external factors.It was also observed that the watertightness of the digital camera section was not maintained due to damage caused by external factors.It was observed that the bending angle was insufficient due to elongation of the angle wire.Lastly, scratches were observed on the following unit components due to external factors: the operational part, operation unit cover, the grip, and on the up-down plate.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus the airway mobilescope battery could not be removed.There was no patient/user harm or injury reported, due to the event.Upon device return and evaluation it was observed that the coating on the light guide serpentine was peeling off, which is a reportable event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record (dhr) found no deviations that could have caused or contributed to the reported event.The dhr confirmed that the subject device was shipped in accordance with the specifications.A definitive root cause for this issue was not established.However, we presume that the defect was caused by stress of repeated use, external factors, or handling.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu) which states the following: ¿5.10 maintenance inspection visually inspect all parts.If necessary, clean them again.Instruments should be treated with lubricants in every reprocessing cycle to maintain their function and to protect them from corrosion and aging.¿ olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|