Manufacturer's investigation conclusion: the report of low power alarms when connected to batteries was confirmed via the log file.The system controller (serial #: (b)(6)) was not returned for analysis; however, the log file was submitted for review with events spanning approximately 14+ days (29 sep 2022 ¿ 13 oct 2022 per timestamp).Decreased relative state of charge (rsoc) and bus voltages were observed on both power cables while connected to the 14v batteries.While connected to battery power, the rsoc/bus voltages fluctuated and dropped below the low voltage thresholds, activating low voltage advisory/hazard alarms; the low voltage hazard alarms bypassed the low voltage advisory alarms multiple times due to the bus voltage dropping to depleted battery voltages.The pump operation was not affected.The pump maintained speeds above the low speed limit throughout the log file.Multiple attempts were made to obtain additional information from the customer regarding the event, including the product return status.However, to date, the system controller (serial #: (b)(6)) was not returned for analysis.Per account information received on 17jan 2023, the product was lost in transit.As a result, the file will be closed with no further action.The exact cause of the reported event was unable to be conclusively determined during this investigation.The device history records were reviewed for the system controller (serial #: (b)(6)) and the system controller was found to pass all manufacturing and quality assurance (qa) specifications before being shipped to the customer.Heartmate ii patient handbook (rev.C) section 5, entitled ¿alarms and troubleshooting¿, and heartmate ii instructions for use (ifu) (rev.C) section 7, entitled ¿alarms and troubleshooting¿, cover all alarms (visual and audible), including the low voltage advisory/hazard alarm conditions, and the actions to take if the alarms cannot be resolved.Heartmate ii patient handbook (rev.C) section 6, entitled "caring for the equipment", describes how to care for and clean all equipment, including the system controller, system controller power cables, 14v batteries, battery clips.Section 10, entitled ¿safety checklists¿, provides checklists to assist the patient in performing routine maintenance of heartmate ii lvad, including inspecting the system controller power cables for dirt, grease, or damage.Heartmate ii patient handbook (rev.C) cautions users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|