Model Number NSLX145C
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 09/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Batch #: v94t5m.Attempts are being made to obtain the following information.To date, no response has been provided.If further details are received at a later date, a supplemental medwatch will be sent: were there any difficulties with device intra-op? how was the bleeding identified? was the site of the bleed identified? was the enseal device used in the procedure at the site of the bleed? was there any surgical intervention necessary to address the bleeding? why does the surgeon believe the post-op bleed may have been caused or contributed by the enseal device? an analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device lot/batch number, and no non-conformances were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that post op to the gastric bypass procedure, the next day, the patient had a bleed by the small bowel and mesentery and they also had a small bleed by the abdominal wall.The patient had to have two blood transfusions and was discharged five days later.The ct scan ruled out that there was leak.The patient is now doing well.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 12/5/2022.Additional information was requested and the following was obtained: were there any difficulties with device intra-op? none that were noticeable how was the bleeding identified? postop drop in hematocrit.Large hematoma found on ct near the jj anastomosis was the site of the bleed identified? believed to be either the omentum, mesentery to the jejunum, or jj anastomosis was the enseal device used in the procedure at the site of the bleed? yes for omentum and mesentery was there any surgical intervention necessary to address the bleeding? no why does the surgeon believe the post-op bleed may have been caused or contributed by the enseal device? enseal device used at site of bleed attempts are being made to obtain the following information.To date, no response has been provided.If further details are received at a later date, a supplemental medwatch will be sent: how was the jejuno-jejunal anastomosis created? which devices were used to create the jejuno-jejunal anastomosis? were they ethicon devices? were there any alleged deficiencies with the ethicon devices that should be included in this complaint?.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL 45 CM SHAFT, CURVED JAW
Common Device Name
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key15711240
Report Number3005075853-2022-07486
Device Sequence Number4972883
Product Code GEI
UDI-Device Identifier10705036015499
UDI-Public10705036015499
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K172580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNSLX145C
Device Catalogue NumberNSLX145C
Device Lot NumberV94T5M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/05/2022
Supplement Date Received by Manufacturer11/09/2022
Initial Report FDA Received Date11/01/2022
Supplement Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-