ETHICON INC. PRLNE BLU 4X24IN 7-0 D/A BV-1 EP; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Model Number EPM8702
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a coronary artery bypass grafting procedure on (b)(6) 2022 and suture was used.The product was used for the blood vessel.The needle came off during use and fell into the body cavity.Although it took time, the needle was retrieved from the body.After the surgeon lost sight of the needle, the inside of the body cavity was checked, but the needle was not found, but then it was found outside the patient's body (it was found on the body).It was not a control release needle.No additional treatment was performed.After that, there was no problem with the patient's condition.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: was the needle retrieved during the same procedure? yes.The surgery was a coronary artery bypass grafting, and the product was used for the blood vessel.After the surgeon lost sight of the needle, the inside of the body cavity was checked, but the needle was not found, but then it was found outside the patient's body.(it was found on the body).No additional treatment was performed.After that, there was no problem with the patient's condition.The following information was requested, but unavailable: were x-rays taken to locate the needle? what measures were taken to retrieve the needle? was there any additional tissue damage as a result of searching for the needle? do you have any photos available for visual analysis? please provide the lot number: this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 4X24IN 7-0 D/A BV-1 EP
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15711403
Report Number2210968-2022-09024
Device Sequence Number12417522
Product Code GAW
UDI-Device Identifier10705031134805
UDI-Public10705031134805
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEPM8702
Device Catalogue NumberEPM8702
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2022
Initial Report FDA Received Date11/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-