• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 10055445
Medical Device Problem Codes Use of Device Problem (1670); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2022 a patient underwent a posterior fixation procedure from l4 to l5.During the procedure while attempting to affix a cross connector the device locking cam appeared to be damaged.The surgeon attempted to remove the engaged driver but it was caught on the driver head and had to be forcibly removed breaking the implant.No patient injury or adverse consequences reported.
 
Additional Manufacturer Narrative
The device was return and photos were provided confirming the alleged complaint.No torque information could be provided and it is unknown of the surgeon utilized hand tightening.Review of the provided photographs found one of the lateral cams had disassociated from the connector with mal positioned rod contact marks additionally cams rotational marks are bent and flared suggesting excessive force.Review of the reported event identified lock damage incurred during attempted placement.Review of all information received indicates excessive force during lock initiation while incomplete rod engagement resulting in damage to the locking mechanism.No additional investigation necessary.Label review "warnings, cautions and precautions - cross connectors are designed specific to the rod diameter and cannot be used on the tapered section of tapered rods.If using cross connectors on tapered rods, only attach them on constant diameter rod sections.Care should be taken to insure that all components are ideally fixated prior to closure." "all implants should be used only with the appropriately designated instrument (reference surgical technique)." should not be scratched or damaged." ".Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk blvd
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk blvd
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk blvd
san diego, CA 92121
MDR Report Key15712001
Report Number2031966-2022-00237
Device Sequence Number14735655
Product Code NKB
UDI-Device Identifier00887517534705
UDI-Public887517534705
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K143684
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10055445
Device Lot NumberN332546
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2022
Initial Report FDA Received Date11/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-