Model Number N/A
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2022
Event Type  malfunction  
Event Description
It was reported at hematology, there were 2 statlock bandages where the hard pvc piece detaches from the adhesive.This report addresses the first device.
 
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of jugq2029 showed one other similar product complaint(s) from this lot number.The complaints for this lot number: (jugq2029) have been reported from facility in belgium.
 
Event Description
It was reported at hematology, there were 2 statlock bandages where the hard pvc piece detaches from the adhesive.This report addresses the first device.
 
Manufacturer Narrative
Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample analysis, applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of detached retainers is confirmed and was determined to be manufacturing related.Two picc plus statlock devices with tricot pads and adjustable posts were returned for evaluation.An initial visual observation showed use residue on one of the returned samples.The retainer of this sample was found to be almost completely detached from its pad.The bottom left corner of the retainer of the other returned sample was also found to be detached from its pad.The pads and the undersides of the retainers of the returned samples were observed to be somewhat tacky.A microscopic observation revealed somewhat poor adhesive coverage on the underside of the detached retainers.Strings of adhesive were observed between the pad and the retainer of both samples, suggesting that the adhesive did not cure correctly.The investigation was forwarded to the manufacturing facility for further evaluation, and bd is working closely with the manufacturing facility to prevent recurrence of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STATLOCK SLIDING POST TRICOT CRESCENT PAD INCLUDES SKIN PREP
Type of Device
STAND, INFUSION
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DAVOL SURGICAL INNOVATIONS -9616067
ave. roberto fierro #6408
parque industrial aeropuerto
cd. juarez, chih s.a. de c.v. 32690
MX   32690
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15712043
MDR Text Key307566418
Report Number3006260740-2022-04233
Device Sequence Number1
Product Code FOX
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberPIC0220CE
Device Lot NumberJUGQ2029
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-