| Catalog Number |
3910-200-030 |
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the anchor broke during the procedure and potentially remains in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke during the procedure and potentially remains in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: anchor broke during surgery.Probable root cause: design: - anchor not designed to withstand tensile forces after insertion.Process: -anchor not manufactured to specification.Application: -excessive force.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
|
| |
|