| Model Number |
J516H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Report initially submitted on 9/15/2022.A response was received from the fda to resubmit the report because ack 3 was not generated and the report was stuck in transmission.The fda gateway prevented transmission of this report.Please reference fda ticket csh-28330 for additional details.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts are being made to obtain the following information. to date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Device return follow up.Events reported via 2210968-2022-07597.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a caesarean section on an unknown date in 2022 and the suture was used.During surgery, when suturing peritoneal adipose tissue, thread is broken during knotting, and the broken part is the suture body.The needle does not fall into the patient's body.There were no adverse patient consequences reported.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 11/30/2022.This report is being submitted pursuant to the provisions of 21 cfr, part 803.Part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6, d9, h3.H3 investigation summary: the product was returned for evaluation.Visual inspection and functional test evaluation were conducted on the returned device.Visual analysis of the returned sample determined that one opened box with seventeen packets and one opened sample that pertains to the product code j516h was returned for analysis with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packets.As per the sampling plan, a visual inspection was performed on the opened sample and thirteen samples, and no defects were found on the packages.The samples were opened, and the swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand and no defects, damage, or suture breakage were noted.A functional test was performed using an instron equipment and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.
|
| |
|
Search Alerts/Recalls
|
|
|