Model Number 1011880J
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a 90% stenosed vein graft in the right posterior descending artery (pda).The 014 ht bmw elite 190cm guidewire (gw) was advanced and used in the procedure; however, after removal without issue, while wiping the gw down for reuse, the gw broke in two pieces.There was no adverse patient effect and no clinically significant delay reported in the procedure.Another bmw gw was used to complete the procedure.No additional information regarding this issue was provided.
 
Additional Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
The reported device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported material separation.Factors that may contribute to a material separation include, but are not limited to, tensile strength, corrosion, excessive force applied to device, interaction with accessory devices, and/or interaction with challenging anatomy.In this case, 50 unused/sterile units from the same batch were returned.A sample size of 19 unused/sterile units were tested and reported no break or material separations.The testing results indicate the reported material separation is likely not related to a product quality issue.It is possible that the guide wire may have become damaged during the procedure due to interaction with the patient¿s 90% stenosed lesion or an accessory device, resulting in the reported material separation; however, this cannot be confirmed.There is no indication of a product quality issue with respect to manufacture, design, or labeling; therefore, no product-related corrective action will be implemented in this case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRES
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15712322
Report Number2024168-2022-11174
Device Sequence Number7251628
Product Code DQX
UDI-Device Identifier08717648144790
UDI-Public08717648144790
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K103101
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model Number1011880J
Device Catalogue Number1011880J
Device Lot Number2083061
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/10/2022
Supplement Date Received by Manufacturer01/12/2023
Initial Report FDA Received Date11/01/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age88 YR
Patient SexMale
Patient Weight79 KG
-
-