BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number L311
Device Problems Pacing Problem (1439); Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2022
Event Type  malfunction  
Manufacturer Narrative
The complaint device was not returned to bsc so product analysis could not be performed.Investigation determined that the minute ventilation (mv) sensor in ingenio devices has the potential to increase the pacing rate to the maximum sensor rate more quickly than expected.As a result, a software update was released in 2013 that reduced the minimum mv response factor settings and re-distributed the remaining mv response factors to provide a consistent change in sensor rate across the range of mv response factors.
 
Event Description
It was reported that this patient with this pacemaker presented to their health care facility for routine follow up and device optimization.The patient is an active hiker and has device feature, minute ventilation (mv) rate response enabled.Upon further review, it was found when the patient is sitting down and not moving, the patient's maximum sensor rate increases to 145 beats per minute.Technical services (ts) was consulted and discussed programming and troubleshooting recommendations.The plan is to perform additional troubleshooting.At this time, the device remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15712332
MDR Text Key306638194
Report Number2124215-2022-44866
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526559228
UDI-Public00802526559228
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/04/2021
Device Model NumberL311
Device Catalogue NumberL311
Device Lot Number413099
Was Device Available for Evaluation? No
Date Manufacturer Received10/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/19/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
-
-