MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
APB-1-1-HX-ES |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that an axium coil prematurely detached.Coil dislodgement was observed in the ic in prime helix1-1 when it was being embolized over the lvis blue, so a spontaneous detachment occurred when an attempt was made to remove it.About 0.5 cm of the proximal edge of the coil was in the headway duo, but an attempt was made to remove it gently while it was still attached to the microcatheter on 4.2 f fubuki.Unfortunately, just before the headway duo entered the fubuki, the coil mass remained at the tip of the fubuki.When the fubuki was removed while carrying out a manual function immediately, the coil was successfully removed.Certainly, again, the headway duo was engaged and the axium prime helix 1x1 was somehow, the placement was withdrawn.The coil was placed in the intended position.Surgical or medical intervention was required to remove dislodged coil.There was no resistance during delivery.The coil was repositioned with 1 cycle of repositioning.The physician attempted to detach the coil. the devices were prepared as indicated in the package insert. patient blood flow speed was normal.No other complications or symptoms were reported. ancillary devices: fubuki guide catheter, headway duo microcatheter, lvis blue.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that there was no impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis of the axium coil (lot# 223938757) found the axium prime pusher actuator interface (ai) and hypotube break indicator (hbi) were intact.No bends or kinks were found with the pusher.The release wire coin was found still against the lumen stop.The pusher shield coil was found intact.The implant coil was detached from the pusher.The detached implant coil was not returned.The shield coil was stretched, and the release wire distal tip was found extending out from the distal end of the pusher for 0.0900mm which is within specification (specification: 0.0508 ¿ 0.127mm).The release wire was removed, and the retainer ring examined.The retainer ring inner diameter (id) was found damaged (indented).Based on the device analysis and reported information, the customer¿s ¿premature detachment¿ report was confirmed.Possible causes of ¿premature detachment¿ include tortuous anatomy, coil is not retracted in a one-to-one motion with the implant pusher during repositioning, pusher rotation, user advances the coil against resistance, damaged pusher, damage to retainer ring, or detachment ball outer diameter (od) below specification.The implant coil was not returned; therefore, an analysis could not be performed, and any contributing factors could not be assessed.Information regarding patient vessel tortuosity was not provided.Tortuous anatomy and detachment ball od below specification could not be ruled out as a potential cause.It was reported there was no resistance during delivery, the pusher was not rotated, and no bends or kinks were found with the pusher.Pusher rotation, user advances the coil against resistance, and damaged pusher were ruled out as potential causes.In this event it is likely the damage found with the retainer ring id and the repositioning of the implant coil contributed to the premature detachment event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|