LUMENIS LTD SLIMLINE EZ; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0642-394-01
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the fiber broke after 2 minutes of firing during the first use.The procedure was completed with this device.There was no patient complication.
 
Additional Manufacturer Narrative
Upon receipt of this fiber at our quality assurance laboratory, this device was thoroughly analyzed.Visual inspection of the fiber body observed no damage; however, the reported fiber break event was confirmed as analysis identified the fiber tip was cracked and had normal degradation.Also, the fiber connector exhibited burn marks on the face.During functional testing, the aiming beam was dim.A review of the device history record (dhr) confirmed that the device met all material, assembly and performance specifications.Based on the information available, a conclusion code of cause traced to component failure was assigned to this investigation.It is likely that an expected or random component failure without any design or manufacturing could have contributed to the fiber break.
 
Event or Problem Description
It was reported that the fiber broke after 2 minutes of firing during the first use.The procedure was completed with this device.There was no patient complication.
 
Additional Manufacturer Narrative
Upon receipt of this fiber at our quality assurance laboratory, this device was thoroughly analyzed.Visual inspection of the fiber body observed no damage; however, the reported fiber break event was confirmed as analysis identified the fiber tip was cracked and had normal degradation.Also, the fiber connector exhibited burn marks on the face.During functional testing, the aiming beam was dim.A review of the device history record (dhr) confirmed that the device met all material, assembly and performance specifications.Based on the information available, a conclusion code of cause traced to component failure was assigned to this investigation.It is likely that an expected or random component failure without any design or manufacturing could have contributed to the fiber break.
 
Event or Problem Description
It was reported that the fiber broke after 2 minutes of firing during the first use.The procedure was completed with this device.There was no patient complication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIMLINE EZ
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD
hakidma st 6
yokneam AZ 20692 04
IS  2069204
Manufacturer (Section G)
MOLEX LLC
18019 n 25th ave
phoenix AZ 85023
Manufacturer Contact
john stoffregen
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key15712545
Report Number2124215-2022-44749
Device Sequence Number4972960
Product Code GEX
UDI-Device Identifier17290109140176
UDI-Public17290109140176
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/18/2023
Device Model Number0642-394-01
Device Catalogue Number0642-394-01
Device Lot Number0003620318
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/23/2022
Supplement Date Received by Manufacturer01/20/2023
09/21/2023
Initial Report FDA Received Date11/01/2022
Supplement Report FDA Received Date02/16/2023
09/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/06/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-