| Model Number |
DWJ023 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was revised after patient complaint and notified physician.Radiographs were taken and the physician stated the implant appeared to have disassociated from baseplate.Patient noticed a difference in the way his shoulder felt after 1 year after being implanted.Radiographs are available pre and post operative, no remarkable circumstances noted to contribute to this event.Patient required revision after physician documented via radiographs that the glenosphere appeared to become disassociated from well-fixed baseplate.Other implants noted for the reverse shoulder arthroplasty were glenoid baseplate, x 2 screws, humeral cup, poly and humeral stem.The explanted implants were requested to be returned at the time of the revision surgery but have not received these back from hospital to date.
|
| |
|
Additional Manufacturer Narrative
|
|
Please note the correction to d9/h3 -product not returned.The reported event could be confirmed the device was not returned for evaluation; however, x-rays were sent for evaluation.No other evidences were provided for investigation.Since data was provided, the opinion of the medical expert was sough and stated as follows: "the x-rays that you¿ve sent indeed confirm the disassembly of the glenosphere and correction of the disassembly after the revision surgery.There are many reasons why this could happen, but without further clinical information and examination of the retrieval this is difficult to say.Most common reason are material interposition at the taper level (blood, debris, gauze), misalignment of the taper during implantation, insufficient impaction during implantation, taper damage during implantation, shoulder instability, disturbed joint mechanics and trauma.For further assessment we would need: early postoperative x-ray of the index procedure, surgical report on both the index and revision surgery and most desirable, retrieval of the explanted glenosphere".More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.H3 other text: device disposition is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was revised after patient complaint and notified physician.Radiographs were taken and the physician stated the implant appeared to have disassociated from baseplate.Patient noticed a difference in the way his shoulder felt after 1 year after being implanted.Radiographs are available pre and post operative, no remarkable circumstances noted to contribute to this event.Patient required revision after physician documented via radiographs that the glenosphere appeared to become disassociated from well-fixed baseplate.Other implants noted for the reverse shoulder arthroplasty were glenoid baseplate, x 2 screws, humeral cup, poly and humeral stem.The explanted implants were requested to be returned at the time of the revision surgery but have not received these back from hospital to date.
|
| |
|