• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. AIRWAY MOBILESCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. AIRWAY MOBILESCOPE Back to Search Results
Model Number MAF-GM
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found watertightness was not maintained due to perforation of the forceps channel, watertightness was not maintained due to holes in the insertion tube, the bending angle was insufficient due to elongation of the angle wire, the image guide bundle was broken, the coating on the image guide coil was peeling off, watertightness of the digital camera was not maintained, and scratches were found on the insertion part corrugated tube.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
The customer reported to olympus, the airway mobilescope had a pinhole in the tip.Inspection and testing of the returned device found the coating was peeling off the flexible tube.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Additional Manufacturer Narrative
Additional information has been received from the customer.This supplemental report is being submitted to provide this information as well as additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely due to stress of repeated use, external factors, or handling.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer provided additional information on 08nov2022: the event was found during reprocessing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRWAY MOBILESCOPE
Common Device Name
AIRWAY MOBILESCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15712618
Report Number9610595-2022-03566
Device Sequence Number20478968
Product Code EOQ
UDI-Device Identifier04953170340369
UDI-Public04953170340369
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K082720
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAF-GM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2022
Supplement Date Received by Manufacturer11/08/2022
Initial Report FDA Received Date11/01/2022
Supplement Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/16/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-