• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: I-SENS RELION PREMIER CLASSIC BLOOD GLUCOSE TEST SYSTEM; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

I-SENS RELION PREMIER CLASSIC BLOOD GLUCOSE TEST SYSTEM; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number GM505UCE
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 08/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the complaint products were not returned, i-sens analyzed the cause of the high readings using a meter of the retained meter and strips of the same lot.As a result of the control solution test, all results were within the standard range, and the cv% was within the acceptance criteria (below 5%) which satisfied the internal standard in i-sens.Therefore, it is judged that the product operated normally.
 
Event or Problem Description
Customer called because he's had the meter for two months and the meter reads 40-50 points off, reading too high.He uses it several times a day.He said that he would read high with the classic, then take insulin, but his sugar would drop, and it would be low.Customer stated he would have symptoms of low blood sugar, such as dizziness and lack of coordination.He said for example, but these aren't the actual numbers, that his reading would be 110 with the classic, but then his actual blood glucose would be 60 when testing with another meter.Customer stores meter and strips in utility drawer in the kitchen.Customer doesn't have trouble getting sample of blood.Customer changes lancets every time he tests.Type 1 diabetic.Customer washes hands with soap and water and allows hands to dry thoroughly.Customer changes lancets most of the time unless he's in a hurry.Customer has not contacted the paramedics, nor did he go to the er.He adjusted his glucose with sugary drinks and food.Walked customer through the memory and had readings of 56 on (b)(6) 2022 at 6:21 pm; 65 on (b)(6) 2022 at 5:52 pm; 72 on (b)(6) 2022 at 5:41 pm; 301 on (b)(6) 2022 at 2:48 am; 46 on (b)(6) 2022 at 6:57 pm.Typical readings range about 150.Customer is not receiving any oxygen therapy.Customer uses fingertip as a testing site.Replaced meter, 25ct strips and sent rl.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELION PREMIER CLASSIC BLOOD GLUCOSE TEST SYSTEM
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
I-SENS
43, banpo-daero 28-gil
seocho-gu
seoul, 06646
KS  06646
Manufacturer (Section G)
I-SENS WONJU FACTORY
94-1, donghwagongdan-ro
munmak-eup
wonju-si, 26365
KS   26365
Manufacturer Contact
hyojin baik
43, banpo-daero 28-gil
seocho-gu
seoul, 06646
KS   06646
MDR Report Key15712793
Report Number3005621222-2022-00020
Device Sequence Number17796974
Product Code NBW
UDI-Device Identifier00681131285742
UDI-Public(01)00681131280624,(01)00681131285742
Combination Product (Y/N)N
Initial Reporter StatePW
Initial Reporter CountryUS
PMA/510(K) Number
K121133
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date04/02/2024
Device Model NumberGM505UCE
Device Catalogue Number784001
Device Lot NumberUM03BBT3B
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2022
Initial Report FDA Received Date11/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/11/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age66 YR
Patient SexMale
Patient Weight73 KG
-
-