EXACTECH, INC. LOGIC CR TIB INSERT STD, SZ 4, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-47-4009
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/19/2019
Type of Reportable Event Serious Injury
Event or Problem Description
As reported via legal, approximately 2 years post op the initial right tka implant, this male patient was revised due to premature polyethylene wear of the tibial insert.Following the revision surgery, the patient "experiences daily pain and discomfort in his right knee which limits his activities of daily living and impacts his quality of life." no additional information.
 
Additional Manufacturer Narrative
Based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of the subsequent revision cannot be conclusively determined; however, it may have been the result of malalignment between the femoral and tibial components, third body wear, patient-related conditions, or any combination of these possibilities, which led to prosthesis wear.
 
Additional Manufacturer Narrative
Section h10: (d4) catalog number: 02-012-47-4009, serial number: (b)(6).Expiration date: 03-jul-2025, unique identifier (udi) #:(b)(4).(d10) concomitant device(s): 4777508 02-010-03-0340 - logic cr femoral cem, right, sz 4 4874892 02-012-45-4040 - lgc tibial fit tray cem sz 4f / 4t 4972565 200-02-38 - three peg patella 38mm (g4) pma/510(k)number: k111400.(h3) the revision reported was likely the result of prosthesis wear as stated in the legal documentation.However, the root cause of the prosthesis cannot be determined as the device was not available for evaluation and images, pre-revision radiographs, and operative notes were not provided.These devices are used for treatment not diagnosis.(h4) device manufacture date: 05-jul-2017.H7 & h9 section h11: the following sections have corrected information: (d1) brand name: logic cr tib insert std, sz 4, 9mm (h6) component code: 734, bearings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC CR TIB INSERT STD, SZ 4, 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key15713549
Report Number1038671-2022-01391
Device Sequence Number4923565
Product Code JWH
UDI-Device Identifier10885862174444
UDI-Public10885862174444
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K111400
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/03/2025
Device Catalogue Number02-012-47-4009
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2022
Supplement Date Received by Manufacturer06/23/2023
Initial Report FDA Received Date11/02/2022
Supplement Report FDA Received Date07/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/05/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0021-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;