MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 80-90% stenosed target lesion was located in the severely tortuous and severely calcified arteriovenous fistula.A 7.0 x80, 75cm gladiator elite balloon catheter was advanced for dilation.However, during eighth inflation for 20 atmospheres for 1 to 3 minutes, the balloon ruptured.The procedure was completed with another of same device.No patient complications were reported and the patient's status is stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 80-90% stenosed target lesion was located in the severely tortuous and severely calcified arteriovenous fistula.A 7.0 x80, 75cm gladiator elite balloon catheter was advanced for dilation.However, during eighth inflation for 20 atmospheres for 1 to 3 minutes, the balloon ruptured.The procedure was completed with another of same device.No patient complications were reported and the patient's status is stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: a gladiator device was received for analysis, the following attributes were examined.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was present in the balloon which is indicative of a leak.The rated burst pressure for this device is 20 atmospheres as per gladiator elite specification.The returned device was attached to an encore inflation unit and positive pressure was applied and liquid was observed to be leaking from a balloon pinhole located approximately 27mm distally of the proximal markerband.A visual examination observed no damage to the tip of the device.A visual and tactile examination of the shaft of the device found no issues.A visual examination found no issues or damage to the markerbands of the device.This concludes the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15713948
Report Number2124215-2022-44385
Device Sequence Number20483393
Product Code LIT
UDI-Device Identifier08714729809838
UDI-Public08714729809838
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/29/2024
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0028113991
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date11/02/2022
Supplement Report FDA Received Date12/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexFemale
Patient Weight62 KG
-
-