| Model Number |
6IC75 |
| Medical Device Problem Codes |
Crack (1135); Installation-Related Problem (2965)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, the inner cannula cracked after hours of being in place.It was also reported that the tracheostomy did not fit correctly.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted one of the tubes present damage (cracks) on the shaft.A dimensional test was performed the inner diameter of the cannulas were within specification.The outer diameter of the cannulas measured within specification.It was reported that the product was cracked.The reported issue was confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: this device is classified as a disposable medical device.Frequent and routine changes of the disposable inner cannula are recommended and should be evaluated by the attending physician.Sterile only if the protective tray and cover are not open, damaged, or broken.Do not re-sterilize.When using an inner cannula, it can only be replaced by a shiley¿ inner cannula of the same size.It is recommended that a spare disposable inner cannula be with the patient at all times.The disposable inner cannula is designed for single use and should not be cleaned or reused.Shiley¿ adult flexible tracheostomy tubes.Patients in the home care environment should be carefully instructed by a home health care provider in the proper use and handling of the tracheostomy tube and accessory products.Federal law (usa) restricts this device to sale by or on the order of a physician.Duration of patient use of the inner cannula should not exceed twenty-nine (29) days because the manufacturer has not substantiated the use of these devices beyond 29-days.Decisions about disposable inner cannula tube changes should be made by the responsible physician or designate using current medical guidelines and judgment.Visually inspect device for signs of damage.If signs of damage, discard and replace.Do not clean or reuse device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|