MEDOS INTERNATIONAL SÃ RL CH EXPEDIUM SPINE SYSTEM SINGLE INNIE SET SCREW 6.35; ORTHOSIS, SPINAL PEDICLE FIXATION
|
Back to Search Results |
|
| Model Number |
179902000 |
| Medical Device Problem Codes |
Device Slipped (1584); Migration (4003)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Date of event: only the event year is known.Procode: kwq, nkb, osh, mnh, kwp.Device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter facility name: (b)(6) hospital.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in japan as follows: it was reported the patient underwent a posterior cervical fusion (th11 to l4) for ais on (b)(6) 2022.After the surgery, the set screws at right l4 were dislocated and loosened.On october 12, 2022, a revision surgery was performed.The set screws and the screw were replaced.Visually, there was no evidence of widening of the screw head or cross-threading of the set screws.No further information is available.This report involves one expedium spine system single innie set screw 6.35.This is report 2 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product code: 179902000.Lot number: 335231.A manufacturing record evaluation was performed for the finished device product code: 179902000.Lot number: 335231.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 02/02/2022 the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the exp 635 ti si setscrew had no signs of issue or defect.A dimensional inspection was performed for the exp 635 ti si setscrew and met specifications.A functional test was performed, the set screw was inserted into the screw head using a stardrive screwdriver, the set screw was able to assemble correctly and lock with no signs of issue.Migration of the device cannot be confirmed since x-ray evidence was not provided.The complaint condition was not able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the observed condition of the exp 635 ti si setscrew would not contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: the following drawings reflecting the current and manufactured revisions were reviewed: expedium 6.35 polyaxial single innie set screw - titanium dwg-1799-02-000 rev.D / rev.D.Dimensional inspection: drawing: specified dimensions: outer diameter of the device (including threads).Measured dimensions: outer diameter of the device (including threads).Device used: mitutoyo digimatic caliper a20276546.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch.
|
| |
|