| Catalog Number |
2R8403 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported a clearlink system non-dehp solution set leaked from an unspecified location.Additionally, the event occurred after the set had been primed with saline.The set was sent to the unit; however, the leak was observed prior to patient contact.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: upon additional information, the leak was observed "at the drip chamber".H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was discarded; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|