Model Number 39815
Medical Device Problem Code Unraveled Material (1664)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on (b)(6) 2022.It was reported that the tip of the guidewire unraveled.A 035/260 (bx/3) magic torque guidewire was advanced into the lesion.However, the tip of the guidewire unraveled while inside the patient.It was noted that this also happened in a separate case using another of same device.In both cases, the guidewire was removed from the patient and the procedure was successfully completed with good blood flow.No patient complications were reported.However, returned device analysis revealed a detached core wire.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned, and during the product analysis it was observed that the coil wire was unraveled and stretched, the core wire was detached/separated at the distal end section, and the distal tip kinked.The device was sent for material analysis and by sem inspection and confirms the fracture of the core wire.The reported event coil wire unraveled material is confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC TORQUE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15715018
Report Number2124215-2022-42022
Device Sequence Number20497195
Product Code DQX
UDI-Device Identifier08714729114628
UDI-Public08714729114628
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K933334
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/07/2025
Device Model Number39815
Device Catalogue Number39815
Device Lot Number0029358427
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/13/2022
Initial Report FDA Received Date11/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
-
-