Model Number 71675383
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Ambulation or Postural Difficulties (2544)
Date of Event 09/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after a trauma procedure on (b)(6) 2022, an intertan 10s 10mm x 18cm 125d broke postoperatively inside the patient.Revision surgery was performed on (b)(6) 2022 to remove the nail and replaced it with a new intertan nail.The surgeon believes it was a fatigue fracture as the patient did not have a fall or any other trauma.The patient had increasing pain in 8 weeks leading up to the fracture of the nail.The current health status of the patient is unknown.
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device did confirm the stated failure mode.The device was found to be fractured into two pieces, rendering the device inoperable.Both of the fractured components were returned.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.The clinical/medical evaluation concluded that based on the information provided, the clinical root cause of the fracture of the intertan nail and pain cannot be determined.The patient¿s weight bearing status and subject activity level, which were not provided, cannot be ruled out as contributing factors to the reported events.The patient impact beyond the pain and revision cannot be determined.However, it was reported that the patient is ¿much happier.¿ no further clinical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management and information for use files revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.According with inspection drawing, the final inspection includes the verification of part configuration per print.Also, material specification shall control the quality and manufacture of standard grade titanium-6 aluminum-4 vanadium alloy.A contribution of the device to the reported event could be corroborated as the device broke and a revision surgery was needed.Factors that could contribute to the reported event include limited range of motion, patient anatomy, postoperative care or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: case-2022-00125237-1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 10S 10MM X 18CM 125D
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15715130
Report Number1020279-2022-04649
Device Sequence Number5096457
Product Code JDS
UDI-Device Identifier00885556105696
UDI-Public00885556105696
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K040212
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71675383
Device Catalogue Number71675383
Device Lot Number21HM04264
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2022
Supplement Date Received by Manufacturer11/28/2022
Initial Report FDA Received Date11/02/2022
Supplement Report FDA Received Date12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexUnknown
-
-