| Model Number |
APB-3-4-3D-ES |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received a report that the coil became pre-maturely detached inside the microcatheter.The catheter was removed and another coil was used.Coil separation/break and premature detachment occurred.There was no surgical or medical intervention required.The pushwire was not bent or broken.There was no friction or difficulty during the delivery.The physician did not reposition the coil.The physician did not attempt to detach the coil.The physician did not rotate the delivery pusher during the procedure.Continuous flush was administered during the procedure.The reported device and any accessory devices were prepared as indicated in the ifu.There were no patient symptoms or further complications were reported as a result of this event. the patient was undergoing surgery for treatment of a saccular, unruptured left pca aneurysm with a max diameter of 4 mm and a 2 mm neck diameter.It was noted the patient's blood flow was normal and vessel tortuosity was moderate.Ancillary devices include an enboy 6f guide catheter, headway17 microcatheter, and a snchro14 guidewire.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis of the axium prime coil (lot# 223625471) found the axium prime pusher actuator interface (ai) and hypotube break indicator (hbi) were intact.No bends or kinks were found with the pusher.The release wire coin was found against the lumen stop.The implant coil was found not attached to the pusher.The implant coil was found to have broken off from the pusher from the coil shell weld.In addition, the implant coil polypropylene filament broke off from the detach element.The implant coil was not returned.Based on the device analysis and reported information, the customer¿s ¿premature detachment¿ report could not be confirmed.However, the customer¿s ¿coil separation/break¿ report was confirmed.Although the axium prime implant coil was found broken, the implant coil did not detach from the pusher.It is likely that the implant coil became stretched subsequently breaking off from the pusher.The headway 17 catheter was found compatible for use with the axium prime coil.The cause for the reported event could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received that the doctor did not attempt to separate.The catheter was not a medtronic product.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|