DEPUY ORTHOPAEDICS INC US PFC SIGMARP CV TB/IN S3 10.0; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT Back to Search Results
Model Number 96-2031
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 10/17/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Clinical adverse event received for poly wear causing instability and increased pain.Event is serious and is considered severe.Event is possibly related to both device and procedure.Date of implant:(b)(6) 2009.Date of revision: unk.Date of event: (b)(6) 2022.(left knee).Treatment: revision; insert was revised.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b1 (product problem), b5, h6 health effect - clinical code.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Operative notes detail the primary left tka that was done on (b)(6) 2009.During the procedure, the surgeon accidentally cut the anterior portion of the it band with an unknown reciprocating saw causing the joint to intraoperatively dislocate.The surgeon was able to repair the it band with sutures, which allowed the joint to reduce without any issues.As a precaution, the surgeon elected to place a mobile bearing revision tray which can be converted over to a hinged if the patient develops instability issues in the future.The manufacturer of the reciprocating saw is unknown and therefore this issue will not be captured in this complaint.Clinical notes then detail the patient returning on (b)(6) 2022 with complaints of pain and instability.Examination of the left knee shows effusion and lateral joint laxity.The radiology report shows significant medial wear of the poly insert compared to previous radiographs.While the clinical notification details a revision of the insert, the revision notes were not provided at this time.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.The information received, will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers, the investigation closed.Should additional information be received, the information will be reviewed.And the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC SIGMARP CV TB/IN S3 10.0
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15715913
Report Number1818910-2022-21901
Device Sequence Number12043860
Product Code NJL
UDI-Device Identifier10603295240747
UDI-Public10603295240747
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96-2031
Device Catalogue Number962031
Device Lot Number2822256
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2022
Supplement Date Received by Manufacturer10/26/2022
11/14/2022
12/02/2022
Initial Report FDA Received Date11/02/2022
Supplement Report FDA Received Date11/09/2022
11/14/2022
12/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/28/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MBT REVISION CEM TIB TRAY SZ 3; MBT TRAY SLEEVE POR M/L 29MM; PFC*SIGMA/OV/DOME PAT 3PEG,38; SIG C/R POROCOAT FEM LT SZ 3
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient SexFemale
Patient Weight127 KG
-
-