Model Number SC45A
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was not function during operation.One could press the firing trigger, but the device was not working.Use the clip along the anvil and take down the jammed anvil.Completed the surgery safely.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Date of event is 2022, event day and month unknown.Captured as (b)(6) 2022.Batch #626a28.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the one sc45a device was returned with no apparent damage, with the closing trigger and anvil in the closed position with tissue between the jaws, the firing mechanism in the return stroke with the knife not fully back.In addition, a tyvek was received along with the device.One gst45w reload was loaded on the device.The knife was manually returned to the home position, the jaws opens as expected and a clip was found lodged in the tissue and the knife, resulting in the firing mechanisms jamming.The cartridge reload was received fully fired and with the deck damaged.In addition, the device was tested for functionality in the articulated position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.The event reported was confirmed and it is related to improper use of the device.To mitigate the potential for staples getting into the cartridge and interfering with the knife path during device firing, prior to reloading the device, rinse the anvil and cartridge jaw in sterile solution and then wipe the anvil and cartridge jaw to clean any formed but unused staples from the device.Additionally, proper care should be taken when placing the device on the tissue to be stapled, to ensure that no hard obstruction such as a clip is included with the tissue inside the jaws.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON FLEX 45 COMPACT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key15716690
Report Number3005075853-2022-07503
Device Sequence Number5126844
Product Code GDW
UDI-Device Identifier10705036002697
UDI-Public10705036002697
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberSC45A
Device Catalogue NumberSC45A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/10/2022
Initial Report FDA Received Date11/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GST WHITE RELOAD, 45MM, 6 ROW
Patient SexUnknown
-
-