Model Number SCPX117MH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 10/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a 48-year-old hispanic female patient underwent a breast surgery with a 350cc mentor cpx4 plus, smooth, medium height tissue expander and experienced a deflation on the left side post-operatively.As a result, the patient underwent explantation on (b)(6) 2022.
 
Additional Manufacturer Narrative
On (b)(6) 2023, mentor completed evaluation of a returned video of the device.Video evaluation summary: upon visual evaluation of the video provided in the complaint, the injection dome seems to be delaminated, causing leakage.As the product involved in this complaint was not received, the device couldn´t be analyzed according to our procedures.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through the mentor¿s quality system.In addition, this complaint was included under the scope of a corrective action (capa) to address this product issue.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR CPX4 PLUS, SMOOTH, MEDIUM HEIGHT
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key15716735
Report Number1645337-2022-13250
Device Sequence Number4923668
Product Code LCJ
UDI-Device Identifier10081317028366
UDI-Public10081317028366
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K182335
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/13/2025
Device Model NumberSCPX117MH
Device Catalogue NumberSCPX117MH
Device Lot Number9642512
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer03/30/2023
Initial Report FDA Received Date11/02/2022
Supplement Report FDA Received Date04/03/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexFemale
Patient EthnicityHispanic
-
-