• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL PROCLAIM¿ DRG IMPLANTABLE PULSE GENERATOR; DRG IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL PROCLAIM¿ DRG IMPLANTABLE PULSE GENERATOR; DRG IPG Back to Search Results
Model Number 3664
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Post Operative Wound Infection (2446)
Event Date 10/10/2022
Event Type  Injury  
Event Description
It was reported that the patient experienced an infection at the ipg site due to a staple being left behind.As a result, surgical intervention was undertaken on (b)(6) 2022 wherein the entire system was explanted to address the issue.The infection has since been resolved.
 
Manufacturer Narrative
Date of event is estimated.Further information requested, but not yet received.
 
Manufacturer Narrative
The device history record was reviewed to ensure proper packaging and sterility.A review of the lot history record identified no manufacturing nonconformities issued to the reported device that would have contributed to this event.Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ DRG IMPLANTABLE PULSE GENERATOR
Type of Device
DRG IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15716892
MDR Text Key302876972
Report Number1627487-2022-05808
Device Sequence Number1
Product Code PMP
UDI-Device Identifier05415067020215
UDI-Public05415067020215
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/07/2024
Device Model Number3664
Device Catalogue Number3664
Device Lot NumberT00001807
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/10/2022
Initial Date FDA Received11/02/2022
Supplement Dates Manufacturer Received01/24/2023
Supplement Dates FDA Received01/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DRG LEAD (X2).
Patient Outcome(s) Other;
Patient SexMale
Patient Weight73 KG
-
-