• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PVC - PORTEX TUBES BLUSELECT; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PVC - PORTEX TUBES BLUSELECT; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/860/080CZ
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2022
Event Type  malfunction  
Manufacturer Narrative
Device evaluation: the device was returned for investigation.A visual inspection was performed.One used decontaminated product was returned for investigation.Under visual inspection it was noticed that pilot balloon was separated from inflation line.There is evident small plastic piece of pilot balloon on inflation line which indicates that bonding between pilot balloon and inflation line was strength enough.Split area was visually inspected under magnification and it seems to be like clear cut by a sharp tool.History of customer complaints was reviewed and any similar customer complaint has been received therefore this customer complaint is considered to be isolated incident with unknown root cause.The cause of the reported problem could not be determined.No lot number was provided; therefore, dhr (device history review) could not be performed.Device expiration date and device manufacturing date are not available.
 
Event Description
It was reported that during the use of the product, the pilot balloon got detached from the inflation line.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PVC - PORTEX TUBES BLUSELECT
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15718175
MDR Text Key307210591
Report Number3012307300-2022-26642
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076158
UDI-Public15019517076158
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/080CZ
Device Catalogue Number101/860/080CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2022
Was the Report Sent to FDA? No
Date Manufacturer Received10/07/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-