MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
3507150BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
09/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: desire to remove breast implants.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 55-year-old female patient underwent an unspecified breast surgery with a mentor memorygel breast implant 150cc gel breast prosthesis that ruptured post implantation.Rupture of the patient¿s right breast prosthesis was observed during a bilateral explantation surgery on (b)(6) 2022.The removed breast prostheses were not replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
On 04-nov-2022, the mentor failure analysis lab received the device for evaluation.On 09-nov-2022, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient¿s right breast prosthesis ruptured.The product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the breast implant was found to be ruptured.In addition, a crease/fold was observed on the edges of the rupture.The evaluation determined that the cause of the rupture is consistent with normal wear.The instructions for use states that ruptures can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following may cause implants to rupture: stressing the implant during implantation or other procedures and weakening it; folding or wrinkling of the implant shell.Breast implants may also wear over time.A second product was received (lot #5844802).No adverse events were reported for this concomitant (contralateral) device.Therefore, no further analysis is required.It should be noted that as part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|