EXACTECH, INC. LOGIC; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
|
Back to Search Results |
|
| Model Number |
CC TIBIAL INSERT SZ 3, 9MM |
| Medical Device Problem Codes |
Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Cyst(s) (1800); Failure of Implant (1924); Pain (1994); Synovitis (2094); Osteolysis (2377); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Aseptic loosening (4883)
|
| Date of Event |
04/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of the subsequent revision cannot be conclusively determined, insufficient information.
|
| |
|
Event or Problem Description
|
|
It was reported via a legal notification, that a 71 yo male patient, initial right knee implanted on (b)(6) 2009 with polyethylene tibial lining manufactured by exactech, inc, suffered personal injuries from that surgery, including but not limited to bone loss and the need for further surgery.No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
D10.Concomitants: tray, tibia trapezoid cementedsize3f/3t, cat# 204-04-33, sn (b)(6); patella, three peg 35mm, cat# 200-02-35, sn (b)(6); extension, femoral andtibialflutedstem 12mm od x11mm l, cat# 204-32-01, sn (b)(6).Component, femur non¿modular cck size 3, cat#210-01-03, sn (b)(6).H3: investigation results - the revision reported was likely the result of wear of the tibial insert and suspected aseptic (non-infected) loosening.The cause and extent of the prosthesis wear cannot be determined because the revised components were not returned for evaluation and no x-rays or photographs were provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, b3, b5, g1, g2, h6- health effect - clinical code, component code, type of investigation, investigation findings, h7 and h11.The revision reported was likely the result of wear of the tibial insert and suspected aseptic (non-infected) loosening.The cause and extent of the prosthesis wear cannot be determined because the revised components were not returned for evaluation and no x-rays or photographs were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial right total knee arthroplasty.Subsequently, it was reported that the patient experienced pain, swelling, osteolysis and prosthesis wear.As a result, the patient underwent a right knee revision approximately 13 years after initial implantation.Patient was last known to be in stable condition.No further information available.
|
| |
|
Search Alerts/Recalls
|
|
|