• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, Back to Search Results
Model Number CC TIBIAL INSERT SZ 3, 9MM
Medical Device Problem Codes Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cyst(s) (1800); Failure of Implant (1924); Pain (1994); Synovitis (2094); Osteolysis (2377); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Aseptic loosening (4883)
Date of Event 04/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of the subsequent revision cannot be conclusively determined, insufficient information.
 
Event or Problem Description
It was reported via a legal notification, that a 71 yo male patient, initial right knee implanted on (b)(6) 2009 with polyethylene tibial lining manufactured by exactech, inc, suffered personal injuries from that surgery, including but not limited to bone loss and the need for further surgery.No additional information.
 
Additional Manufacturer Narrative
D10.Concomitants: tray, tibia trapezoid cementedsize3f/3t, cat# 204-04-33, sn (b)(6); patella, three peg 35mm, cat# 200-02-35, sn (b)(6); extension, femoral andtibialflutedstem 12mm od x11mm l, cat# 204-32-01, sn (b)(6).Component, femur non¿modular cck size 3, cat#210-01-03, sn (b)(6).H3: investigation results - the revision reported was likely the result of wear of the tibial insert and suspected aseptic (non-infected) loosening.The cause and extent of the prosthesis wear cannot be determined because the revised components were not returned for evaluation and no x-rays or photographs were provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, b3, b5, g1, g2, h6- health effect - clinical code, component code, type of investigation, investigation findings, h7 and h11.The revision reported was likely the result of wear of the tibial insert and suspected aseptic (non-infected) loosening.The cause and extent of the prosthesis wear cannot be determined because the revised components were not returned for evaluation and no x-rays or photographs were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that the patient underwent an initial right total knee arthroplasty.Subsequently, it was reported that the patient experienced pain, swelling, osteolysis and prosthesis wear.As a result, the patient underwent a right knee revision approximately 13 years after initial implantation.Patient was last known to be in stable condition.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
ashley harris
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key15718528
Report Number1038671-2022-01395
Device Sequence Number4897337
Product Code JWH
UDI-Device Identifier10885862054241
UDI-Public10885862054241
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K954208
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/26/2014
Device Model NumberCC TIBIAL INSERT SZ 3, 9MM
Device Catalogue Number208-23-09
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer12/27/2023
03/16/2026
Initial Report FDA Received Date11/02/2022
Supplement Report FDA Received Date01/03/2024
03/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/27/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0019-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient Age71 YR
Patient SexMale
-
-