• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-030
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2022, a 30mm amplatzer septal occluder was selected for procedure.It was noted that during procedure the occluder exhibited a deformation during deployment.The decision was made to abort the implant procedure.There were no adverse patient consequences reported.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2022, a 30mm amplatzer septal occluder was selected for procedure.It was noted that during procedure the occluder exhibited a deformation during deployment.The decision was made to aborted the implant procedure.There were no adverse patient consequences reported.Subsequent to the previously file report, additional information was received that the delivery system used when attempting to implant the 30mm amplatzer septal occluder was a non-abbott 10f delivery system.The 30mm amplatzer septal occluder was not damaged or mishandled in any during device preparation.The deformed occluder was confirmed/visualized while inside the patient via fluoroscopy examination.The deformed 30mm amplatzer septal occluder was never released from the delivery cable while inside the patient.There was interaction with cardiac structures during deployment and no kink or angulation noted in the delivery system.The implant procedure was aborted due to the lack of another 30mm amplatzer septal occluder or similar sized device.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects reported.
 
Additional Manufacturer Narrative
The reported event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Based on the information received, the cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15718870
Report Number2135147-2022-01899
Device Sequence Number20720681
Product Code MLV
UDI-Device Identifier05415067019332
UDI-Public05415067019332
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9-ASD-030
Device Catalogue Number9-ASD-030
Device Lot Number8462842
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/12/2022
Supplement Date Received by Manufacturer01/12/2023
Initial Report FDA Received Date11/02/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-