| Model Number |
MU-651RA |
| Medical Device Problem Codes |
Power Problem (3010); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer reported that the bsm was monitoring a patient and it randomly shut off and would not power back on.They were monitoring a patient, but no injury or harm occurred.They put the patient on a different bsm to continue monitoring the patient.Nka technical support troubleshot the issue with no resolution.The bsm would not power on.The customer put a charged battery into the unit and the bsm powered back on, but only with the charged battery.Once the battery was removed, the unit shut off again.The unit was sent in for repair and during evaluation the reported problem was verified.The malfunctioning parts were replaced and the unit was tested and operates to the manufacturer's specifications.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported that the bsm was monitoring a patient and it randomly shut off and would not power back on.They were monitoring a patient, but no injury or harm occurred.
|
| |
|
Event or Problem Description
|
|
The customer reported that the bedside monitor (bsm) randomly shut down on its own while monitoring a patient and would not power back on.The bsm would only power on when a charged battery was inserted.No patient injury or harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Details of complaint: the customer reported that the bedside monitor (bsm) randomly shut down on its own while monitoring a patient and would not power back on.The bsm would only power on when a charged battery was inserted.No patient harm or injury was reported.Investigation summary: the reported device was sent in for repair.During the evaluation of the returned device, nihon kohden repair center (nk rc) was able to duplicate the reported issue of the unit not powering on when plugged into ac power.They were able to power on the unit using battery power.The power unit was replaced to resolve the issue.A possible cause is likely related to hardware failure of the bsm's power unit.Hardware failure could occur because of physical damage, heat damage, or electrical damage.Physical damage could occur due to impacts with objects and other surfaces.Heat damage could occur due to improper maintenance or device placement.Electrical damage could occur during a power outage or power surge.The unit was installed in 09/2013 with no history of servicing.The following fields contain no information (ni), as attempts to obtain information were made, but not provided: attempt # 1: 10/24/2022 emailed the customer via microsoft outlook for all information in the ni list above: no reply was received.Attempt # 2: 10/26/2022 emailed the customer via microsoft outlook for all information in the ni list above: the customer responded, stating they would not be able to provide the requested information.Additional information: b4 date of this report.G3 date received by manufacturer.G6 type of report.H2 if follow-up, what type? h6 event problem and evaluation codes.
|
| |
|
Event or Problem Description
|
|
The customer reported that the bedside monitor (bsm) randomly shut down on its own while monitoring a patient and would not power back on.The bsm would only power on when a charged battery was inserted.No patient injury or harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates** corrected information: d1 brand name: corrected the brand name from bsm-6501a to life scope tr.D4 additional device information / model #: corrected the model # from bsm-6501a to mu-651ra.D4 additional device information / primary unique device identifier (udi) #: corrected the udi # to include the pi information.This is a correction to the suspect medical device involved in the reported event, specifically the unique device identifier (udi) information in section d of the fda form 3500a.
|
| |
|
Search Alerts/Recalls
|