MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 7MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 7MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 7MM
Medical Device Problem Codes Fracture (1260); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Failure of Implant (1924); Periprosthetic fracture (4902)
Date of Event 09/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported, female patient had a periprosthetic fracture of the stem.Patient was revised to hrp.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.X-rays and photos received.The device is not available for evaluation due to hospital policy.
 
Additional Manufacturer Narrative
Section b5: additional information received indicates that patient's bone was broken on distal part of the stem.Section d10: concomitant products: equinoxe reverse 38mm glenosphere (cat#: 320-01-38 / serial#: (b)(6).Equinoxe reverse tray adapter plate tray +0 (cat#: 320-10-00 / serial#: (b)(6).Eq rev glenoid plate (cat#: 320-15-01 / serial#: (b)(6).Eq rev locking screw (cat#: 320-15-05 / serial#: (b)(6).Eq reverse torque defining screw kit (cat#: 320-20-00 / serial#: (b)(6).Eq rev compress screw lck cap kit, 4.5 x 22mm (cat#: 320-20-22 / serial#: (b)(6).Eq rev compress screw lck cap kit, 4.5 x 22mm (cat#: 320-20-22 / serial#: (b)(6).Eq rev compress screw lck cap kit, 4.5 x 26mm (cat#: 320-20-26 / serial#: (b)(6).Eq rev compress screw lck cap kit, 4.5 x 30mm (cat#: 320-20-30 / serial#: (b)(6).145-deg pe 38mm hum liner +0 (cat#: 320-38-00 / serial#: (b)(6).(h3) as reported, female patient had a periprosthetic fracture of the stem.Patient was revised to hrp.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.X-rays and photos received.The device is not available for evaluation due to hospital policy.Upon review of the available information, there is no evidence that this is a device related problem and there is no allegation against the device.Implantation of a total joint could result pain, infection, or loosening of total joint hardware.The most likely cause of the reported event was a malunion of a fracture which was likely caused by over-reaming during implantation.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b5, d2a, d2b, d8, h6: clinical code, medical device problem code, component code, investigation conclusions.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
X-ray and photos received did not pertain to this patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 7MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key15720381
Report Number1038671-2022-01373
Device Sequence Number2995765
Product Code KWS
UDI-Device Identifier10885862079299
UDI-Public10885862079299
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K042021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/14/2019
Device Model NumberEQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 7MM
Device Catalogue Number300-01-07
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2022
Supplement Date Received by Manufacturer04/17/2023
11/06/2025
Initial Report FDA Received Date11/03/2022
Supplement Report FDA Received Date05/09/2023
11/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/15/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-