Model Number STRIP, TF WGN 50CTMG/DL 956970
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Meter and test strips were not returned for evaluation.Added most likely underlying root cause onto case as the complaint product was not returned for investigation.Most likely underlying root cause: mlc-006: passed open expiration date.Note: manufacturer contacted customer in several follow-up calls to ensure the initial concern is resolved - unable to establish contact with customer at this time.
 
Event or Problem Description
Consumer reported complaint for lo blood glucose test results.Wife is calling on behalf of the customer.Wife stated she had tried to test her blood glucose using her husband's meter.Wife stated that she turned the true focus meter on, inserted the test strip with the blood on it and saw lo, and that the lo remains on the screen.Customer's wife stated that her husband had obtained lo recently.The customer¿s expected blood glucose test result range was not disclosed.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call, a back to back blood test was not performed by the customer.The test strip lot manufacturer¿s expiration date is 08/24/2023 and test strips were opened a year ago (expired).Product storage was not disclosed.The meter memory was not reviewed for previous test result history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUE FOCUS
Common Device Name
SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key15722161
Report Number1000113657-2022-00579
Device Sequence Number20634263
Product Code NBW
UDI-Device Identifier00021292008956
UDI-Public(01)00021292008956
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/24/2023
Device Model NumberSTRIP, TF WGN 50CTMG/DL 956970
Device Lot NumberNY1010
Was Device Available for Evaluation? No
Distributor Facility Aware Date10/07/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/07/2022
Initial Report FDA Received Date11/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-