Model Number SA6AT6
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Disturbances (2140)
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient experienced blurring left eye, impedes driving.The clinical reason for explant was lens off target.The iol was exchanged for another lens model 3 months following the initial implant procedure.Additional information has been requested there are two medical device reports associated with this patient.This report is associated with the 2 of 2 files.Follow up information received as this is not a bilateral file - but an explanted lens for replacement iol used in case (b)(4).
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information has been received and stated that the iol shifted, at which point the surgeon felt like a capsular tension ring (ctr) would help stabilize the iol.However, while attempting to place the ctr the patient was found to have quite a bit of scarring/adhesions to the bag and it was felt that trying to place a ctr was too risky; at that point the surgeon opted to explant the iol and implant a standard iol with option for lasik to correct his astigmatism at a later date.Currently the patient is doing well and has had no adverse complications.There are two medical device reports associated with this patient.This report is associated with the 2 of 2 files.
 
Additional Manufacturer Narrative
Correction information provided in d.10.Additional information provided in h.3.,h.6.And h.10.The account indicated the use of a qualified associated cartridge and handpiece.The viscoelastic indicated is not qualified for use with this lens with any qualified lens/cartridge/handpiece combinations.The product investigation could not identify a root cause for the reported complaint.The explanted lens was not returned.Information in the file indicated that the replacement lens for the company lens 9.0 diopter was a non-company +11.00 diopter lens.The manufacturer internal reference number is:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF TORIC ASPHERIC UV ABSORBING SINGLEPIECE IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15722358
Report Number1119421-2022-02340
Device Sequence Number7251964
Product Code HQL
UDI-Device Identifier00380652360838
UDI-Public00380652360838
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model NumberSA6AT6
Device Lot Number12639831
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/07/2022
Supplement Date Received by Manufacturer11/02/2022
12/07/2022
Initial Report FDA Received Date11/03/2022
Supplement Report FDA Received Date12/01/2022
01/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LI61AO, +11.00,4861304050; MONARCH II CARTRIDGE B; MONARCH III INJECTOR; NUVISC
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-