| Model Number |
368346 |
| Medical Device Problem Code |
Extracardiac stimulation (1463)
|
| Health Effect - Clinical Code |
Undesired Nerve Stimulation (1980)
|
| Date of Event |
10/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This lead was explanted due to phrenic stimulation.Should additional information become available, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt, the lead under complaint was found cut 13 cm distal to the is-1 connector pin (into two segments).All fragments were received for analysis.An extraction aid was found on the distal fragment.The analysis revealed further cuts into the insulation 60 cm proximal to the lead tip.These cuts most likely resulted from the extraction procedure.Further analysis of the lead fragments did not show any deviations which might have led to the reported phrenic stimulation.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|
| |
|