ETHICON INC. VCL RAP UD 27IN 3-0 S/A PS-1 PRM; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number VR935
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/21/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent a vaginal delivery on (b)(6) 2022 and suture was used.During the procedure, the suture pulled off the needle when sewing the wound in the perineal incision.The needle was not found at that time.The next day, it was found in the photo that the needle was at the muscle layer of the perineal wound.Under spinal anesthesia, the needle was removed from the perineal wound, the suture was replaced, and the perineal wound was sutured again.Additional information was requested.
 
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.(b)(4).To date it has been reported that the device will not be returned.If the device or further details are received as a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please provide the patient's demographic information including age, weight, bmi at the time of index procedure on what tissue was the suture used? what was the tissue condition (normal, thin, calcified, fragile, diseased)? how was the suture placed (interrupted or continuous)? how was the suture originally tied (multiple knots, square knot, etc.)? what instruments were used in this procedure to handle the suture? did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status? if applicable, will product be returned? if so, please provide the return date and tracking information.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, and the following was obtained: please provide the patient's demographic information including age, weight, bmi at the time of index procedure -unk on what tissue was the suture used? - muscular layer of perineum what was the tissue condition (normal, thin, calcified, fragile, diseased)? -unk how was the suture placed (interrupted or continuous)? -continuous how was the suture originally tied (multiple knots, square knot, etc.)? -unk what instruments were used in this procedure to handle the suture? -needle holder did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? -no.What is the physician¿s opinion as to the etiology of or contributing factors to this event? -unk what is the patient's current status?: good.If applicable, will product be returned? if so, please provide the return date and tracking information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL RAP UD 27IN 3-0 S/A PS-1 PRM
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-BRAZIL
rodovia presidente dutra
km 154
sao paolo 12240 -908
BR   12240-908
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15723080
Report Number2210968-2022-09095
Device Sequence Number2996354
Product Code GAM
UDI-Device Identifier10705031134799
UDI-Public10705031134799
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K033746
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVR935
Device Catalogue NumberVR935
Device Lot NumberAR6397
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2022
Supplement Date Received by Manufacturer11/09/2022
Initial Report FDA Received Date11/03/2022
Supplement Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age27 YR
Patient SexFemale
-
-